Tél: +86-13333857585 E-mail: enquête@otywell-tector.com

À propos Contact |

Analyseur de gaz infrarouge de type d'aspiration de pompe G2000 - Otywell | Fabricant d'analyseurs de gaz

Analyseur de gaz/

Analyseur de gaz infrarouge de type d'aspiration de pompe G2000

  • Continuous online detection
  • Sound and light alarm indication
  • Chinese LCD display
  • Built-in high-power carbon film self-priming pump
  • AC220V power supply
  • Sensor self-check function
  • Alarm value can be set
  • Secondary flow controllable
  • Modular infrared sensor
  • Alarm record query function
  • Convenient desktop support
  • Power-on function self-check
  • Convenient precision fine-tuning function
  • Convenient one-button pump start function

Mobile / whatsapp: +86-13333857585Email: enquête@otywell-tector.com

  • Product Details

Description

G2000 Pump suction type infrared gas analyzer adopts built-in power carbon film self-priming pump infrared detection principle, instrument is imported infrared sensor and advanced digital processing technology, Can carry out continuous on-line analysis of the gas (CO2, CO, CH4 and NOX, SO2, etc). Chinese LCD display, sound and light alarm, long life, high sensitivity, fast response, can be set up alarm value, boot self-checking function, with functions of alarm record. used in petroleum, chimique, metallurgical and other industrial and agriculture, medical and health departments and so on.

Fonctions fonctionnalités

  • Continuous detecting on-line
  • liquid crystal display (Écran LCD)in chinese
  • Sound and light alarm indicator
  • A built-in power carbon film self-priming pump
  • Sensor detection
  • Set the alarm value
  • Secondary flow control
  • Aluminum chassis
  • Modular infrared sensor>
  • Record alarm function
  • Convenient desktop support type
  • Boot self-check function
  • Convenient precision trimming function/li>
  • A key to switch on pump conveniently
Principe de détectionElectrochemical, infrarouge, laser, etc
Detect gasFlammable gas, toxic and harmful gases, etc
Méthode d'échantillonnagePump suction type
Test rangeBased on gas type
Differentiation rate1%Lel;1 Ppm / 0.1 Ppm.0.01
Measurement error+ 3%FS
Status indicationliquid crystal indication
Debug mode key operation
Temps de réponse< 20s
Recovery time< 20s
Température ambiante-10 ~ 50
Flux de travail500mL/min
Degré de protectionIP65
AlimentationAC220V
SourceDC7.2v 2800mA
WeightNNo more than 2000g
Limite de pression86kPa ~ 106kPa
Taille31 * 27.2 * 27.2 (mm)
PoidsÀ propos 4.5 kilos